The Hygiena EnSURE Touch is a handheld luminometer and monitoring system. It reads Hygiena test devices, stores the results, and syncs them to cloud software so a verification programme produces records rather than a notebook.
It has a 5 inch touch screen, wireless sync, and cloud-based reporting through SureTrend.
What ATP testing actually tells you
ATP — adenosine triphosphate — is the energy molecule present in all living cells and in residues derived from them. An ATP test does not identify organisms and it is not a microbiological count. What it measures is organic residue: food soil, body fluids, biofilm and the living cells within them.
That is exactly the right question for verifying cleaning. A surface that has been cleaned properly has had the residue removed; a surface that looks clean and still returns a high reading has not. The result arrives in seconds, while the cleaner is still on site and can do something about it.
What it is not: a test for pathogens, a substitute for microbiological sampling, or a measure of disinfection. A sanitiser can kill organisms and leave their ATP behind. Use ATP to verify that cleaning removed the soil — which is the step disinfection depends on.
What it reads
The EnSURE Touch is compatible with three families of Hygiena test device:
AquaSnap Free — water ATP (not currently a line we carry)
Indicator organism tests — the MicroSnap range (Coliform, E. coli, Enterobacteriaceae, Total), which give same-day results and sit between an ATP reading and a full culture. Not lines we currently carry — ask us.
Enzyme tests — including SpotCheck Plus for glucose and lactose. Hygiena also list CrossCheck for acid phosphatase and ZymoSnap ALP for alkaline phosphatase; neither is currently a line we carry.
Why the data side matters more than the meter
Any luminometer will give you a number. What separates a monitoring system from a gadget is whether those numbers become a record you can defend.
The EnSURE Touch syncs results over WiFi to SureTrend Cloud, where readings from multiple sites sit in one dashboard, with user access controls and the ability to trend results over time.
That matters for three practical reasons:
A trend beats a reading. One high result is an argument. Twelve months of results showing a surface consistently passing, and the one week it did not, is evidence.
Results stop depending on who took them. A test recorded against a named location by a named user at a known time is auditable. A number written on a clipboard is not.
You can act on patterns rather than incidents. Recurring failures at the same location point at the procedure or the access, not at the cleaner.
Setting it up properly
Define your test points before you start. The value of the system is comparison over time, and that requires the same locations, sampled the same way, every time. Naming them properly at set-up saves months of unusable data.
Set your own pass and fail limits from your own baseline. There is no universal RLU figure that means clean. Sample surfaces you know have been properly cleaned, and derive your limits from those.
Agree the sampling technique — area, pressure, direction — and train to it. Most of the variation in ATP programmes comes from swabbing technique, not from the instrument.
Decide where in the cleaning cycle you sample. Sanitiser still wet on a surface can suppress the reading. Consistency matters more than which point you choose.
Choose the SystemSURE Plus if you need readings and will keep your own records. Choose the EnSURE Touch if the records themselves are part of what you are being asked to produce.
Checking the Meter Itself
Hygiena state the EnSURE Touch runs a sensor check at power-up and has a built-in tilt monitor, which warns when the instrument is held at an angle that could affect a reading. For a documented check against a known light source, use the Hygiena CalCheck — Hygiena state it verifies calibration in under 60 seconds and the result syncs to SureTrend Cloud with your test data.
Specifications
Brand
Hygiena
Hygiena part number
ETOUCH (ASY0460)
SKU
HG-ETOUCH
Screen
5 inch shatter-resistant touch screen
Size
Approximately 3 × 7 × 1 in (76 × 178 × 25 mm) (Hygiena)
Sensor
Photodiode
Result time
10 seconds (Hygiena)
Sensitivity
Below 1 femtomole ATP (Hygiena)
Connectivity
Wi-Fi with SSL encryption; syncs to SureTrend Cloud
Self-checks
Sensor check at power-up; tilt monitor
Remote support
Built-in TeamViewer screen sharing (Hygiena)
Carry
Shoulder strap and hand strap
Battery, weight, warranty
Not published on Hygiena's product page — ask us
Frequently asked questions
Does it come with test devices?
No. The swabs and enzyme tests are consumables ordered separately.
What RLU number is a pass?
Yours to set, from your own baseline. Any figure quoted without reference to your surfaces and your soil is not evidence about your site.
Does it need internet access to work?
It reads and stores results on the device; the cloud sync is what moves them to SureTrend over WiFi. Talk to us about how your site is set up if connectivity is a constraint.
Can it identify bacteria?
ATP tests measure organic residue, not organisms. The MicroSnap indicator organism tests are the devices that address that, and they run on the same meter.
Can I use non-Hygiena swabs in it?
No. The meter and the devices are a matched system.
That is Hygiena's position: the EnSURE Touch is validated with Hygiena test devices. Note that Anaeron, the maker of CleanTrust swabs, lists its CleanTrust ATP Surface Swab as compatible with the EnSURE Touch — that is Anaeron's claim, not Hygiena's. If you mix swab brands, readings are not comparable, so set a separate baseline for each.
How do I know the meter itself is reading correctly?
Use a Cal Check — a reusable calibration verification device. We carry one.
Is the battery life, weight or warranty published?
Not on Hygiena's product page. Ask us and we will get the current figures from Hygiena before you order.
Specifying a cleaning verification system? Tell us what you have to demonstrate, to whom, and across how many sites — we will help you work out whether you need the cloud reporting or just the readings.
Payment & Security
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Estimate shipping
Refund Policy
Refund Policy
At The Moisture Meter Company (TMMC), a division of Carpet Cleaners Warehouse, we are committed to complying with the Australian Consumer Law (ACL) and providing fair outcomes for our customers.
Australian Consumer Law Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement, repair, or refund for a major failure and compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if they fail to be of acceptable quality and the failure does not amount to a major failure.
Damaged, Faulty or Incorrect Goods If you receive a product that is damaged in transit, faulty, defective, or incorrectly supplied, please contact us as soon as possible.
To assist with assessment, we may request photographs, serial numbers, proof of purchase, or other information relevant to the claim.
Where a product is confirmed to be faulty, damaged, incorrectly supplied, or otherwise covered under Australian Consumer Law, TMMC will provide an appropriate remedy, which may include: - Repair - Replacement - Store credit - Refund
TMMC will cover reasonable return shipping costs for approved warranty, fault, damage, or incorrect supply claims.
Change of Mind Returns TMMC may, at its sole discretion, accept returns for change of mind within 14 days of delivery, provided that: - The return is approved by TMMC before being sent back. - The product is unused and in original condition. - The product is returned in its original packaging. - All accessories, manuals, and included items are returned. - The product is not a special order, made-to-order, clearance, consumable, or non-returnable item.
Approved change of mind returns may be subject to a restocking fee and the customer is responsible for all return shipping costs.
Original freight charges are non-refundable unless required under Australian Consumer Law.
Non-Returnable Items Unless required under Australian Consumer Law, the following items cannot be returned for change of mind: - Special order products - Made-to-order products - Hygiena swabs, test reagents and other consumables - Probes, electrodes and pins that have been used - Products specifically marked as non-returnable - Clearance or end-of-line items
Return Process Before returning any product, customers must obtain return authorisation from TMMC.
Please contact us using one of the methods below: - Phone: 1800 68 68 69 - Email: info@moisturemeters.com.au
Returns sent without prior approval may be refused or delayed.
Refund Processing Approved refunds will be processed to the original payment method used for the purchase unless otherwise agreed.
Refunds are generally processed after the returned product has been received, inspected, and approved.
Processing times may vary depending on the payment provider.
Warranty Claims Many products sold by TMMC are also covered by manufacturer warranties. These are in addition to your rights under the Australian Consumer Law.
Warranty claims may require inspection, testing, repair, replacement, or manufacturer assessment before a final outcome is determined.
Equipment must be returned to TMMC in Queensland for warranty assessment. The customer pays freight to return the item; if a covered fault is confirmed, TMMC reimburses that freight and pays return freight. If the fault is not covered, we will provide a written quote before any work is done.
Business Customers Nothing in this policy excludes, restricts, or modifies any rights available under Australian Consumer Law.
Where permitted by law, remedies available for goods supplied for business use may be limited in accordance with applicable legislation.
Contact Us The Moisture Meter Company Phone: 1800 68 68 69 Email: info@moisturemeters.com.au